A 65-nm CMOS receiver tailored for breast cancer diagnostic imaging is demonstrated for the first time. The receiver shows 31-dB conversion gain, NF< 8.6dB , P1dB> -28dBm, IIP3 >-12dBm and IIP2>22dBm over a band from 1.75 to 15 GHz. A programmable injection-locked divider generates quadrature LO signals with a I/Q phase error <1.5deg over three octaves without requiring any calibration or tuning. The receiver shows a flicker noise corner as low as 40 Hz, achieving a dynamic range of 106 dB.

A 65-nm CMOS 1.75-15 GHz Stepped Frequency Radar Receiver for Early Diagnosis of Breast Cancer

BASSI, MATTEO;CARUSO, MICHELE;BEVILACQUA, ANDREA;NEVIANI, ANDREA
2013

Abstract

A 65-nm CMOS receiver tailored for breast cancer diagnostic imaging is demonstrated for the first time. The receiver shows 31-dB conversion gain, NF< 8.6dB , P1dB> -28dBm, IIP3 >-12dBm and IIP2>22dBm over a band from 1.75 to 15 GHz. A programmable injection-locked divider generates quadrature LO signals with a I/Q phase error <1.5deg over three octaves without requiring any calibration or tuning. The receiver shows a flicker noise corner as low as 40 Hz, achieving a dynamic range of 106 dB.