Background: Urinary catecholamine metabolites are well-established biomarkers for neuroblastoma (NB). Homovanillic acid (HVA) and vanillylmandelic acid (VMA) are the most frequently measured metabolites within SIOPEN-Catecholamine Working Group laboratories. Here, we evaluated the performance of a new LC-MS/MS in vitro diagnostic (IVD) kit for HVA and VMA to facilitate inter-laboratory harmonization. Methods: HVA and VMA and their deuterated internal standards were analyzed with a commercial method, on a ThermoFisher Quantiva LC-MS/MS. Validation was performed first using internal quality control and external quality assessment (IQC and EQA) samples. Next by clinical validation on 120 samples, previously tested by HPLC-ECD. Finally, 36 samples were exchanged between SIOPEN reference laboratories and analyzed by three methods. Results: Using QCs and EQA the method was validated in a wide calibration range (4.61-830 mu mol/L for HVA and 4.44-800 mu mol/L for VMA). Intra-day CVs (n = 5) were 7 and 8 % for HVA and 5 and 6 % for VMA for QC low and QC high, respectively; Inter-day CV% were 7 and 3 % for HVA and 2 and 7 % for VMA at QC low and QC high, respectively. Its application to 120 clinical samples confirmed a high diagnostic accuracy. The inter-laboratory quality control assessment showed interchangeable results (p = 0,73 and p = 0.15 for HVA and VMA, respectively).
Validation of an LC-MS/MS method for urinary homovanillic and vanillylmandelic ACIDS and application to the diagnosis of neuroblastoma
Rossi, LucillaInvestigation
;Mancin, FabrizioProject Administration
;
2025
Abstract
Background: Urinary catecholamine metabolites are well-established biomarkers for neuroblastoma (NB). Homovanillic acid (HVA) and vanillylmandelic acid (VMA) are the most frequently measured metabolites within SIOPEN-Catecholamine Working Group laboratories. Here, we evaluated the performance of a new LC-MS/MS in vitro diagnostic (IVD) kit for HVA and VMA to facilitate inter-laboratory harmonization. Methods: HVA and VMA and their deuterated internal standards were analyzed with a commercial method, on a ThermoFisher Quantiva LC-MS/MS. Validation was performed first using internal quality control and external quality assessment (IQC and EQA) samples. Next by clinical validation on 120 samples, previously tested by HPLC-ECD. Finally, 36 samples were exchanged between SIOPEN reference laboratories and analyzed by three methods. Results: Using QCs and EQA the method was validated in a wide calibration range (4.61-830 mu mol/L for HVA and 4.44-800 mu mol/L for VMA). Intra-day CVs (n = 5) were 7 and 8 % for HVA and 5 and 6 % for VMA for QC low and QC high, respectively; Inter-day CV% were 7 and 3 % for HVA and 2 and 7 % for VMA at QC low and QC high, respectively. Its application to 120 clinical samples confirmed a high diagnostic accuracy. The inter-laboratory quality control assessment showed interchangeable results (p = 0,73 and p = 0.15 for HVA and VMA, respectively).| File | Dimensione | Formato | |
|---|---|---|---|
|
1-s2.0-S2667145X25000148-main.pdf
accesso aperto
Tipologia:
Published (Publisher's Version of Record)
Licenza:
Creative commons
Dimensione
4.59 MB
Formato
Adobe PDF
|
4.59 MB | Adobe PDF | Visualizza/Apri |
Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.