Background: We compared the performance of 21 different assays performed by the Wantai Wan200+ (Wantai BioPharm, Beijing, China) with respect to other methods in use at the University Hospital of Padova (AOPD), Italy. Methods: The plasma (P) or serum (S) of 5027 leftover samples, collected from May to Sept 2023, was either analyzed or frozen at -20 degrees C. Beckman DXI800 (DXI), Roche Cobas 8000 e801 (RC), Snibe Maglumi 4000 plus (SM), DiaSorin Liaison XL (DL) and Binding Site Optilite (BS) equipment were used at the AOPD. P-procalcitonin (PCT), DXI; P-Troponin I (TnI), DXI; S-CA125, DXI; S-free PSA (f-PSA), DXI; S-total PSA (t-PSA), DXI; S-IL6, SM; P-Troponin T (TnT), RC; P-NT-proBNP, RC; P-Neuron-Specific Enolase (NSE), RC; S-CA15-3, DL; S-CA19-9, DL; S-AFP, DL; and S-CEA, DL were tested in fresh samples. P-Myoglobin (Myo), DXI; P-Cyfra21-1, RC; S-beta 2 microglobulin (B2MIC), BS; S-HE4, SM; S-PGI, SM; S-PGII, SM; S-CA72-4, SM; and S-CA50, SM were analyzed in frozen and thawed samples. Bland-Altman (BA), Passing-Bablok (PB) and Cohen's Kappa (CKa) metrics were used as statistics. Results: An excellent comparability profile was found for 11 analytes. For example, the t-PSA CKa was 0.94 (95%CI: 0.90 to 0.98), and the PB slope and intercept were 1.02 (95%CI: 0.99 to 1.03) and 0.02 (95%CI: 0.01 to 0.03), respectively; the BA bias was 2.25 (95%CI: -0.43 to 4.93). Ten tested measurands demonstrated a suboptimal comparability profile. Biological variation in EFLM (EuBIVAS) performance specifications was evaluated to assess the clinical relevance of measured biases. Conclusions: Evaluation of the Wantai Wan200+'s performance suggests that between-method differences did not exceed the calculated bias. Metrological traceability may influence the comparisons obtained for some measurands.
Evaluation of the Comparability of Wantai Wan200+ Instrument with Routine Laboratory Assays for 21 Different Analytes
Talli, Ilaria;Padoan, Andrea;Cosma, Chiara;Galozzi, Paola;Basso, Daniela;
2024
Abstract
Background: We compared the performance of 21 different assays performed by the Wantai Wan200+ (Wantai BioPharm, Beijing, China) with respect to other methods in use at the University Hospital of Padova (AOPD), Italy. Methods: The plasma (P) or serum (S) of 5027 leftover samples, collected from May to Sept 2023, was either analyzed or frozen at -20 degrees C. Beckman DXI800 (DXI), Roche Cobas 8000 e801 (RC), Snibe Maglumi 4000 plus (SM), DiaSorin Liaison XL (DL) and Binding Site Optilite (BS) equipment were used at the AOPD. P-procalcitonin (PCT), DXI; P-Troponin I (TnI), DXI; S-CA125, DXI; S-free PSA (f-PSA), DXI; S-total PSA (t-PSA), DXI; S-IL6, SM; P-Troponin T (TnT), RC; P-NT-proBNP, RC; P-Neuron-Specific Enolase (NSE), RC; S-CA15-3, DL; S-CA19-9, DL; S-AFP, DL; and S-CEA, DL were tested in fresh samples. P-Myoglobin (Myo), DXI; P-Cyfra21-1, RC; S-beta 2 microglobulin (B2MIC), BS; S-HE4, SM; S-PGI, SM; S-PGII, SM; S-CA72-4, SM; and S-CA50, SM were analyzed in frozen and thawed samples. Bland-Altman (BA), Passing-Bablok (PB) and Cohen's Kappa (CKa) metrics were used as statistics. Results: An excellent comparability profile was found for 11 analytes. For example, the t-PSA CKa was 0.94 (95%CI: 0.90 to 0.98), and the PB slope and intercept were 1.02 (95%CI: 0.99 to 1.03) and 0.02 (95%CI: 0.01 to 0.03), respectively; the BA bias was 2.25 (95%CI: -0.43 to 4.93). Ten tested measurands demonstrated a suboptimal comparability profile. Biological variation in EFLM (EuBIVAS) performance specifications was evaluated to assess the clinical relevance of measured biases. Conclusions: Evaluation of the Wantai Wan200+'s performance suggests that between-method differences did not exceed the calculated bias. Metrological traceability may influence the comparisons obtained for some measurands.Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.