Closed-loop stimulation (CLS; BIOTRONIK SE & Co. KG, Berlin, Germany) is a rate-responsive algorithm that analyzes intracardiac impedance trends using a standard lead placed in the right ventricle. It is unknown whether CLS could perform adequately with His bundle (HB) lead placement, as contractility dynamics may be attenuated in this region compared to the right ventricle apex.We performed hand-grip, mental, and bicycle exercise tests in a patient with brady atrial fibrillation and permanent HB pacing. The CLS algorithm responded with an appropriate heart rate to mental and physical tests.
Closed loop stimulation with His bundle lead placement
Giacopelli, Daniele;
2022
Abstract
Closed-loop stimulation (CLS; BIOTRONIK SE & Co. KG, Berlin, Germany) is a rate-responsive algorithm that analyzes intracardiac impedance trends using a standard lead placed in the right ventricle. It is unknown whether CLS could perform adequately with His bundle (HB) lead placement, as contractility dynamics may be attenuated in this region compared to the right ventricle apex.We performed hand-grip, mental, and bicycle exercise tests in a patient with brady atrial fibrillation and permanent HB pacing. The CLS algorithm responded with an appropriate heart rate to mental and physical tests.File in questo prodotto: