INTRODUCTION: A critical point for using blood purification during sepsis may be the potential interaction with antimicrobial therapy, the mainstay of sepsis treatment. The aim of our study was to investigate the vancomycin removal during hemoperfusion (HP) using HA380 cartridge.METHODS: This is an experimental study, in which 500 mL of solution was circulated in a closed-circuit (blood flow of 250 mL/min) simulating HP ran using HA380. Vancomycin was added to reach a through concentration or a very high concentration to evaluate the removal ratio (RR) during 120 min of HP. Comparison between blood-crystalloid solution and balanced solution was performed by using Kruskal-Wallis test. The kinetics of vancomycin removal and the adsorption isotherm were evaluated.RESULTS: We found a complete removal of vancomycin at baseline through concentration of 23.0 ± 7.4 mg/L. Using extremely high concentration (baseline 777.0 ± 62.2 mg/L), RR was 90.1 ± 0.6% at 5 min and 99.2 ± 0.6% at 120 min. No difference in terms of RR was found between blood-crystalloid mixture and balanced solution. The kinetics of the vancomycin reduction followed an exponential decay. Repeated boluses (total amount of 2,000 mg) resulted in cumulative adsorption of 1,919.4 mg with RR of 96.6 ± 1.4%, regardless of the amount injected (100 vs. 500 mg). Vancomycin adsorption onto HA380 followed the Langmuir isotherm model.CONCLUSIONS: A considerable amount of vancomycin was rapidly removed during in vitro HP with HA380. Clinical studies are needed to determine whether this may lead to underdosing. Drug therapeutic monitoring is highly recommended when using HA380 for blood purification in patients receiving vancomycin.
Vancomycin Adsorption During in vitro Model of Hemoperfusion with HA380 Cartridge
Godi, Ilaria;Golino, Gianlorenzo;Sandini, Alessandra;de Cal, Massimo;Navalesi, Paolo;Ronco, Claudio
2021
Abstract
INTRODUCTION: A critical point for using blood purification during sepsis may be the potential interaction with antimicrobial therapy, the mainstay of sepsis treatment. The aim of our study was to investigate the vancomycin removal during hemoperfusion (HP) using HA380 cartridge.METHODS: This is an experimental study, in which 500 mL of solution was circulated in a closed-circuit (blood flow of 250 mL/min) simulating HP ran using HA380. Vancomycin was added to reach a through concentration or a very high concentration to evaluate the removal ratio (RR) during 120 min of HP. Comparison between blood-crystalloid solution and balanced solution was performed by using Kruskal-Wallis test. The kinetics of vancomycin removal and the adsorption isotherm were evaluated.RESULTS: We found a complete removal of vancomycin at baseline through concentration of 23.0 ± 7.4 mg/L. Using extremely high concentration (baseline 777.0 ± 62.2 mg/L), RR was 90.1 ± 0.6% at 5 min and 99.2 ± 0.6% at 120 min. No difference in terms of RR was found between blood-crystalloid mixture and balanced solution. The kinetics of the vancomycin reduction followed an exponential decay. Repeated boluses (total amount of 2,000 mg) resulted in cumulative adsorption of 1,919.4 mg with RR of 96.6 ± 1.4%, regardless of the amount injected (100 vs. 500 mg). Vancomycin adsorption onto HA380 followed the Langmuir isotherm model.CONCLUSIONS: A considerable amount of vancomycin was rapidly removed during in vitro HP with HA380. Clinical studies are needed to determine whether this may lead to underdosing. Drug therapeutic monitoring is highly recommended when using HA380 for blood purification in patients receiving vancomycin.Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.