The purpose of this study was to evaluate the interindividual and intraindividual variability of slow motorized pull-through lower esophageal sphincter (LES) manometry compared to standard station pull-through LES manometry to measure LES overall length, abdominal length, and pressure and to report normal values for the slow motorized pull-through method. The slow motorized pull-through had significantly smaller coefficient of variation, indicating closer agreement between different examiners in analyzing a given tracing. The correlation coefficients for each parameter in normal subjects and symptomatic patients was significantly higher when using slow motorized pull-through for both patients and normal subjects for all three parameters. The 5th percentile of normal values obtained from 41 volunteers for LES overall length, abdominal length, and pressure was 2.7 cm, 1.4 cm, 5.1 mm Hg, respectively. The results indicate that the slow motorized pull-through method is more reproducible than the standard station pull-through method both between different observers and when the same examiner measures the same tracing on two different occasions.

Manometry of the lower esophageal sphincter: Inter- and intraindividual variability of slow motorized pull-through versus station pull-through manometry

Costantini M.;
2003

Abstract

The purpose of this study was to evaluate the interindividual and intraindividual variability of slow motorized pull-through lower esophageal sphincter (LES) manometry compared to standard station pull-through LES manometry to measure LES overall length, abdominal length, and pressure and to report normal values for the slow motorized pull-through method. The slow motorized pull-through had significantly smaller coefficient of variation, indicating closer agreement between different examiners in analyzing a given tracing. The correlation coefficients for each parameter in normal subjects and symptomatic patients was significantly higher when using slow motorized pull-through for both patients and normal subjects for all three parameters. The 5th percentile of normal values obtained from 41 volunteers for LES overall length, abdominal length, and pressure was 2.7 cm, 1.4 cm, 5.1 mm Hg, respectively. The results indicate that the slow motorized pull-through method is more reproducible than the standard station pull-through method both between different observers and when the same examiner measures the same tracing on two different occasions.
2003
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11577/3333830
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 14
  • ???jsp.display-item.citation.isi??? 12
  • OpenAlex ND
social impact