This article focuses on the static structural analysis of a tibial intramedullary nail, carried out by the finite element method. The investigated subject is concerned with the issue of frictional contacts that, generating a non-linear problem, play an important role in numerical analyses. Experimental tests were also developed on tibial intramedullary nail specimens. The experimental results, in terms of force–displacement, were used to calibrate and validate the finite element analysis, by tuning the contact parameters available in the software.

Experimentally validated structural finite element method analysis of a tibial intramedullary nail: Optimal choice of the contact settings

Croccolo, D;Fini, S;Robusto, F
2019

Abstract

This article focuses on the static structural analysis of a tibial intramedullary nail, carried out by the finite element method. The investigated subject is concerned with the issue of frictional contacts that, generating a non-linear problem, play an important role in numerical analyses. Experimental tests were also developed on tibial intramedullary nail specimens. The experimental results, in terms of force–displacement, were used to calibrate and validate the finite element analysis, by tuning the contact parameters available in the software.
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11577/3285145
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 2
  • ???jsp.display-item.citation.isi??? 2
  • OpenAlex ND
social impact