Purpose The aim of this paper is to investigate the limits of LET monitoring of therapeutic carbon ion beams with miniaturized microdosimetric detectors. Methods Four different miniaturized microdosimeters have been used at the 62 MeV/u 12C beam of INFN Southern National Laboratory (LNS) of Catania for this purpose, i.e. a mini-TEPC and a GEM-microdosimeter, both filled with propane gas, and a silicon and a diamond microdosimeter. The (dose-mean lineal energy) values, measured at different depths in a PMMA phantom, have been compared with (dose-mean LET) values in water, calculated at the same water-equivalent depth with a Monte Carlo simulation setup based on the GEANT4 toolkit. Results In these first measurements, no detector was found to be significantly better than the others as a LET monitor. The relative standard deviation has been assessed to be 13% for all the detectors. On average, the ratio between and values is 0.9 ± 0.3, spanning from 0.73 ± 0.08 (in the proximal edge and Bragg peak region) to 1.1 ± 0.3 at the distal edge. Conclusions All the four microdosimeters are able to monitor the dose-mean LET with the 11% precision up to the distal edge. In the distal edge region, the ratio of to changes. Such variability is possibly due to a dependence of the detector response on depth, since the particle mean-path length inside the detectors can vary, especially in the distal edge region.
Miniaturized microdosimeters as LET monitors: First comparison of calculated and experimental data performed at the 62 MeV/u 12C beam of INFN-LNS with four different detectors
Conte, V.;Selva, A.;De Nardo, L.;
2018
Abstract
Purpose The aim of this paper is to investigate the limits of LET monitoring of therapeutic carbon ion beams with miniaturized microdosimetric detectors. Methods Four different miniaturized microdosimeters have been used at the 62 MeV/u 12C beam of INFN Southern National Laboratory (LNS) of Catania for this purpose, i.e. a mini-TEPC and a GEM-microdosimeter, both filled with propane gas, and a silicon and a diamond microdosimeter. The (dose-mean lineal energy) values, measured at different depths in a PMMA phantom, have been compared with (dose-mean LET) values in water, calculated at the same water-equivalent depth with a Monte Carlo simulation setup based on the GEANT4 toolkit. Results In these first measurements, no detector was found to be significantly better than the others as a LET monitor. The relative standard deviation has been assessed to be 13% for all the detectors. On average, the ratio between and values is 0.9 ± 0.3, spanning from 0.73 ± 0.08 (in the proximal edge and Bragg peak region) to 1.1 ± 0.3 at the distal edge. Conclusions All the four microdosimeters are able to monitor the dose-mean LET with the 11% precision up to the distal edge. In the distal edge region, the ratio of to changes. Such variability is possibly due to a dependence of the detector response on depth, since the particle mean-path length inside the detectors can vary, especially in the distal edge region.File | Dimensione | Formato | |
---|---|---|---|
Physica medica 2018.pdf
accesso aperto
Descrizione: articolo principale
Tipologia:
Published (publisher's version)
Licenza:
Creative commons
Dimensione
1.29 MB
Formato
Adobe PDF
|
1.29 MB | Adobe PDF | Visualizza/Apri |
Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.