In this work we present the characterization of a large area 200 mm ×200 mm EJ-444 scintillation detector to be used for monitoring gross alpha and beta activity in tap water plants. Specific tests were performed in order to determine the best setup to read-out the light from the detector side. The possibility to stack many detectors and get a compact device with total active area of the order of 1 m2 has been explored. Alpha/beta discrimination, efficiency and homogeneity tests were carried out with alpha and beta sources. Background from ambient radioactivity was measured as well. Alpha/beta real-time monitoring in drinking water is a goal of the EU project TAWARA_RTM.

Characterization of a Large Area ZnS(Ag) Detector for Gross Alpha and Beta Activity Measurements in Tap Water Plants

CESTER, DAVIDE;LUNARDON, MARCELLO;MORETTO, SANDRA;STEVANATO, LUCA;VIESTI, GIUSEPPE
2016

Abstract

In this work we present the characterization of a large area 200 mm ×200 mm EJ-444 scintillation detector to be used for monitoring gross alpha and beta activity in tap water plants. Specific tests were performed in order to determine the best setup to read-out the light from the detector side. The possibility to stack many detectors and get a compact device with total active area of the order of 1 m2 has been explored. Alpha/beta discrimination, efficiency and homogeneity tests were carried out with alpha and beta sources. Background from ambient radioactivity was measured as well. Alpha/beta real-time monitoring in drinking water is a goal of the EU project TAWARA_RTM.
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11577/3187817
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 13
  • ???jsp.display-item.citation.isi??? 9
  • OpenAlex ND
social impact