Abstract BACKGROUND: The present study was conducted to evaluate the analytical performance and the organizational aspects of Capillarys 2 Flex Piercing system (CFP) respect to agarose electrophoresis and HPLC methods in hemoglobinopathies screening. METHODS: The measurement of imprecision in HbA 2 and HbF quantification was verified on HbA 2 CFP control and on three samples; 74 whole blood samples were used to evaluate migration time imprecision of hemoglobin variants S, C and E (HbS, HbC, and HbE); to compare methods, 451 samples were tested on CFP and HPLC; reference values were verified as value distribution in 160 blood donors and at ROC curve analysis on 449 samples from routine analysis. RESULTS: Imprecision: the analytical CV % s ranged from 1.25 to 3.9 at HbA 2 quantification, the CV % was 3.78 at HbF quantification; the running time imprecision for HbS and HbC and HbE ranged from 0.20 to 0.69 % . Method comparison: at regression analysis findings were HbA 2: CFP=1.21×HPLC–0.64, HbF: CFP=1.31×HPLC−0.75, HbS: CFP=1.10×HPLC−3.24. Reference values: the HbA 2 95th percentile range was 2.5–2.8; HbF was undetectable in 154 out 160 samples tested; at ROC curve analysis the best combination of sensitivity and diagnostic efficiency was obtained using 2.2 and 3.0, as reference values, for HbA 2 and 1.1 as the upper reference limit for HbF. Organizational aspects: with respect to the procedures currently implemented in our laboratory CFP requires 2 h less time and obviates the need for some manual steps. CONCLUSIONS: The quantification, reproducibility and diagnostic efficiency provided by CFP in identification and quantification of hemoglobins appear accurate. In addition, the use of primary tubes allows improved safety, and the avoidance of some manual steps, that prolong working time and are a source of possible errors.

Identification and quantification of hemoglobins in whole blood: the analytical and organizational aspects of Capillarys 2 Flex Piercing compared with agarose electrophoresis and HPLC methods.

VARAGNOLO, MARIA CRISTINA;ZANINOTTO, MARTINA;PLEBANI, MARIO
2013

Abstract

Abstract BACKGROUND: The present study was conducted to evaluate the analytical performance and the organizational aspects of Capillarys 2 Flex Piercing system (CFP) respect to agarose electrophoresis and HPLC methods in hemoglobinopathies screening. METHODS: The measurement of imprecision in HbA 2 and HbF quantification was verified on HbA 2 CFP control and on three samples; 74 whole blood samples were used to evaluate migration time imprecision of hemoglobin variants S, C and E (HbS, HbC, and HbE); to compare methods, 451 samples were tested on CFP and HPLC; reference values were verified as value distribution in 160 blood donors and at ROC curve analysis on 449 samples from routine analysis. RESULTS: Imprecision: the analytical CV % s ranged from 1.25 to 3.9 at HbA 2 quantification, the CV % was 3.78 at HbF quantification; the running time imprecision for HbS and HbC and HbE ranged from 0.20 to 0.69 % . Method comparison: at regression analysis findings were HbA 2: CFP=1.21×HPLC–0.64, HbF: CFP=1.31×HPLC−0.75, HbS: CFP=1.10×HPLC−3.24. Reference values: the HbA 2 95th percentile range was 2.5–2.8; HbF was undetectable in 154 out 160 samples tested; at ROC curve analysis the best combination of sensitivity and diagnostic efficiency was obtained using 2.2 and 3.0, as reference values, for HbA 2 and 1.1 as the upper reference limit for HbF. Organizational aspects: with respect to the procedures currently implemented in our laboratory CFP requires 2 h less time and obviates the need for some manual steps. CONCLUSIONS: The quantification, reproducibility and diagnostic efficiency provided by CFP in identification and quantification of hemoglobins appear accurate. In addition, the use of primary tubes allows improved safety, and the avoidance of some manual steps, that prolong working time and are a source of possible errors.
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11577/3157361
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 9
  • ???jsp.display-item.citation.isi??? 9
  • OpenAlex ND
social impact