Using a fast digitizer, the neutron–gamma discrimination capability of the new liquid scintillator EJ-309 is compared with that obtained using standard EJ-301. Moreover the capability of both the scintillation detectors to identify a weak neutron source in a high gamma-ray background is demonstrated. The probability of neutron detection is PD=95% at 95% confidence level for a gamma-ray background corresponding to a dose rate of 100 uSv/h.
Neutron detection in a high gamma-ray background with EJ-301 and EJ-309 liquid scintillators
STEVANATO, LUCA;CESTER, DAVIDE;VIESTI, GIUSEPPE
2012
Abstract
Using a fast digitizer, the neutron–gamma discrimination capability of the new liquid scintillator EJ-309 is compared with that obtained using standard EJ-301. Moreover the capability of both the scintillation detectors to identify a weak neutron source in a high gamma-ray background is demonstrated. The probability of neutron detection is PD=95% at 95% confidence level for a gamma-ray background corresponding to a dose rate of 100 uSv/h.File in questo prodotto:
Non ci sono file associati a questo prodotto.