We evaluated performance of the TEST 1 (SIRE Analytical Systems, Udine, Italy), a fully automated analyzer for the measurement of the erythrocyte sedimentation rate (ESR). Intra-assay reproducibility was satisfactory for a wide range of ESR values, whereas there was a significant decrease in ESR data when the samples were stored at 4°C for up to 24 hours. We compared TEST 1 with the Westergren ESR method approved by the International Council for Standardization in Haematology (ICSH) and with the Diesse Ves-Matic analyzer. Linear regression analysis comparing the TEST 1 and the ICSH reference method yielded satisfactory correlations for K3 EDTA- and sodium citrate-anticoagulated samples. Bland-Airman analysis showed no evidence of a systematic bias between the TEST 1 and the reference method. A close correlation was found between the TEST 1 system and the Diesse Ves-Matic analyzer despite a significant positive systematic bias. Reference values for men and women were analyzed according to nonparametric statistics. The TEST 1 was easy to use, had a satisfactory operative praticability, required minimal maintenance, and reduced contact with potential biohazards. This system enables the determination of ESR with any common standard-sized tube; the use of samples anticoagulated with K3 EDTA can widely reduce the workload in clinical laboratories.

The TEST 1 automated system: A new method for measuring the erythrocyte sedimentation rate

PLEBANI, MARIO;SANZARI, MARIA COLOMBA;BERNARDI, DANIELA;
1998

Abstract

We evaluated performance of the TEST 1 (SIRE Analytical Systems, Udine, Italy), a fully automated analyzer for the measurement of the erythrocyte sedimentation rate (ESR). Intra-assay reproducibility was satisfactory for a wide range of ESR values, whereas there was a significant decrease in ESR data when the samples were stored at 4°C for up to 24 hours. We compared TEST 1 with the Westergren ESR method approved by the International Council for Standardization in Haematology (ICSH) and with the Diesse Ves-Matic analyzer. Linear regression analysis comparing the TEST 1 and the ICSH reference method yielded satisfactory correlations for K3 EDTA- and sodium citrate-anticoagulated samples. Bland-Airman analysis showed no evidence of a systematic bias between the TEST 1 and the reference method. A close correlation was found between the TEST 1 system and the Diesse Ves-Matic analyzer despite a significant positive systematic bias. Reference values for men and women were analyzed according to nonparametric statistics. The TEST 1 was easy to use, had a satisfactory operative praticability, required minimal maintenance, and reduced contact with potential biohazards. This system enables the determination of ESR with any common standard-sized tube; the use of samples anticoagulated with K3 EDTA can widely reduce the workload in clinical laboratories.
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11577/2508494
Citazioni
  • ???jsp.display-item.citation.pmc??? 10
  • Scopus 62
  • ???jsp.display-item.citation.isi??? 57
  • OpenAlex ND
social impact