Before the installation in the OPERA experiment, bakelite Resistive Plate Chambers are subjected to a chain of quality tests performed at the Gran Sasso Laboratories. Mechanical properties are first investigated in a fully automatized station, measuring the gas leakage and checking the gluing of the internal spacers. Then electrical properties are tested: current vs. voltage characteristics are measured both with pure argon and with gas mixture filling; moreover the short term behavior of the current is monitored. Finally the response uniformity is tested, measuring the efficiency on the entire detector surface at cosmic rays fluxes. In this paper we present the test chain and the results on a large set of chambers.

The quality control tests for the RPCs of the OPERA experiment

BRUGNERA, RICCARDO;GARFAGNINI, ALBERTO;A. Longhin;
2004

Abstract

Before the installation in the OPERA experiment, bakelite Resistive Plate Chambers are subjected to a chain of quality tests performed at the Gran Sasso Laboratories. Mechanical properties are first investigated in a fully automatized station, measuring the gas leakage and checking the gluing of the internal spacers. Then electrical properties are tested: current vs. voltage characteristics are measured both with pure argon and with gas mixture filling; moreover the short term behavior of the current is monitored. Finally the response uniformity is tested, measuring the efficiency on the entire detector surface at cosmic rays fluxes. In this paper we present the test chain and the results on a large set of chambers.
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11577/2500362
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 8
  • ???jsp.display-item.citation.isi??? 10
social impact