A 68-year-old man underwent reoperation because of severe incompetence of a mitral valve xenograft, 96 months following implantation. Gross examination of the device showed yellow spots on the cusps, suggesting lipid infiltration, and a torn commissure. Plasmatic cholesterol and lipiprotein levels were normal. X-ray examination of the explants showed no calcific deposits. Histologic and electron microscopic studies disclosed massive accumulation of lipid clefts and droplets, predominantly at the level of the tear; focal loss and detachment of the endothelial lining and scanty porcine fibroblasts and collagen bundles with preserved periodicity were also noted. Primary disruption of porcine valvular bioprostheses without significant calcifications or collagen breakdown is uncommon. In the present case, lipid accumulation was the main determinant of tissue failure
Cusp disruption by massive lipid infiltration. A rare cause of porcine valve dysfunction.
VALENTE, MARIALUISA;THIENE, GAETANO
1982
Abstract
A 68-year-old man underwent reoperation because of severe incompetence of a mitral valve xenograft, 96 months following implantation. Gross examination of the device showed yellow spots on the cusps, suggesting lipid infiltration, and a torn commissure. Plasmatic cholesterol and lipiprotein levels were normal. X-ray examination of the explants showed no calcific deposits. Histologic and electron microscopic studies disclosed massive accumulation of lipid clefts and droplets, predominantly at the level of the tear; focal loss and detachment of the endothelial lining and scanty porcine fibroblasts and collagen bundles with preserved periodicity were also noted. Primary disruption of porcine valvular bioprostheses without significant calcifications or collagen breakdown is uncommon. In the present case, lipid accumulation was the main determinant of tissue failurePubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.