We describe the complete calibration of a ruby laser-based multipoint Thomson scattering system, performed only by rotational Raman scattering from H2 and the measurement of the absolute Raman cross section of some rotational transitions suitable for such calibrations. The absolute calibration factors (for ne measurements by Thomson scattering) have been measured by observing the Stokes J = 1 → 3 rotational transition at Δv =−587 cm−1(λ = 723.8 nm), with a ≅4.5% accuracy. The relative calibration (for Te measurements) has also been performed by Raman scattering, observing the same line in all the spectral channels. The results agree to within ≅ 10% with those obtained by a tungsten lamp and a lock-in detection technique. From a comparison of the results of the two methods, systematic errors on the measured Te and ne of less than ˜10% (for Te up to 200 eV) are inferred. The absolute cross section of the rotational Raman transitions at Av = −587 cm−1(λ = 723.8 nm), Δv = −354 cm−1(λ = 711...
Complete calibration of a Thomson scattering spectrometer system by rotational Raman scattering in H2
GIUDICOTTI, LEONARDO
1987
Abstract
We describe the complete calibration of a ruby laser-based multipoint Thomson scattering system, performed only by rotational Raman scattering from H2 and the measurement of the absolute Raman cross section of some rotational transitions suitable for such calibrations. The absolute calibration factors (for ne measurements by Thomson scattering) have been measured by observing the Stokes J = 1 → 3 rotational transition at Δv =−587 cm−1(λ = 723.8 nm), with a ≅4.5% accuracy. The relative calibration (for Te measurements) has also been performed by Raman scattering, observing the same line in all the spectral channels. The results agree to within ≅ 10% with those obtained by a tungsten lamp and a lock-in detection technique. From a comparison of the results of the two methods, systematic errors on the measured Te and ne of less than ˜10% (for Te up to 200 eV) are inferred. The absolute cross section of the rotational Raman transitions at Av = −587 cm−1(λ = 723.8 nm), Δv = −354 cm−1(λ = 711...Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.