The calibration and monitoring system constructed for the HARP experiment scintillator-based time of flight system is described. It is based on a Nd-Yag laser with passive Q-switch and active/passive mode-locking, with a custom made laser light injection system based on a bundle of IR monomode optical fibers. A novel ultrafast InGaAs MSM photodiode, with 30 ps risetime, has been used for the laser pulse timing. The first results from the 2001-2002 data taking are presented, showing that drifts-in timing down to about 70 ps can be traced.

The laser calibration system of the HARP TOF

GIBIN, DANIELE;
2003

Abstract

The calibration and monitoring system constructed for the HARP experiment scintillator-based time of flight system is described. It is based on a Nd-Yag laser with passive Q-switch and active/passive mode-locking, with a custom made laser light injection system based on a bundle of IR monomode optical fibers. A novel ultrafast InGaAs MSM photodiode, with 30 ps risetime, has been used for the laser pulse timing. The first results from the 2001-2002 data taking are presented, showing that drifts-in timing down to about 70 ps can be traced.
2003
8th Topical Seminar on Innovative Particle and Radiation Detectors, Siena, Italy, 21-24 Oct 2002.